We specialize in producing high-quality components and products for a variety of industries. Our team of experienced engineers and technicians work together to ensure that every product we manufacture meets or exceeds our clients' expectations. We pride ourselves on our attention to detail and commitment to delivering exceptional customer service.
We offer a wide range of services including prototyping, product design, and manufacturing. Our state-of-the-art facility and skilled team allow us to provide our clients with custom solutions that meet their unique needs.
Leitech's Inch or Metric Calibration Masters make it easy to reset the specified calibration length on Standard or Hi-Resolution COMBI Gages. Depth calibration should be checked / set on a periodic basis and is required when a new GO member is installed.
1. Dial in the required thread pitch on the Master to establish the specified calibration length.
2. Loosen the two locking set screws (A) on the COMBI Handle.
3. Place the gage in the appropriate size hole in the Master.
4. Turn the NOT GO end in either direction to set the COMBI Gage measuring scale on the sleeve to either the .400 line for the inch model, or the 10,0 mm line on the metric.
5. Re-tighten the locking screws.
Your COMBI Gage is now set at the correct calibration value for accurate functional thread depth measurement. This fast and foolproof method eliminates the need for a vernier caliper, depth micrometer, or gage blocks.
Note: Common size nuts available from customer inventories are tightened against the removal sleeve (similar to a jam nut) to remove the taperlock member from the Leitech handle.
The gaging members on the Leitech COMBI gage are of a special taperlock design which can be easily changed with a removal sleeve that can be purchased from Leitech U.S. and a nut of the correct size available from customer inventories.
1. Slide the COMBI removal sleeve (Green for the GO & Red for the NOT GO) over the gage member to be removed.
2. Thread the nut (supplied by the user) on to the gage member, and tighten it against the removal sleeve (similar to a jam nut) until the gage member is pulled loose from the handle.
3. Insert the new thread member and give it a light tap against a soft surface to seat the tapered portion of the gage member in the handle.
4. Loosen the two locking screws "A" on the COMBI handle to permit depth calibration of the measuring sleeve. Now, the user can turn the threaded inner portion of the handle body which extends or retracts the face of the GO gage member in relation to the face of the measuring sleeve. Note: The measuring sleeve is set a few threadsback from the face of the GO member to permit easier entry into the product thread.
5. Place the gage in a Leitech Calibration Master as described on the reverse side. As an alternative, a gage block stack corresponding to the calibration length (X) in the chart below may be used. In this procedure, the face of the measuring sleeve should be resting on the block stack and the face of the GO member should be against a reference plane,such as a surface plate. A depth micrometer or a vernier caliper may be used in a similar fashion.
6. Turn the NOT GO end in either direction and line up the measuring scale on the sleeve with the .400 line on the inch model, or the 10,00 mm line for the metric, at the intersection of the measuring sleeve and the handle body.
7. Re-tighten the locking screws finger tight.
Your COMBI gage is now set at the correct calibration value for accurate functional thread depth measurement. The user can discriminate thread depths to an accuracy of approximately 1/2 division or .012 in. (0,25 mm).